The Scenario
This worked example walks through a simplified static bending-stress calculation for a hip implant stem, following the same mechanics-of-materials workflow used in this site's Implant Load & Stress Calculator: an 80 kg patient, a joint reaction force factor of 2.8× body weight (a representative single-leg-stance value), a stem neck offset of 40 mm, and a solid circular stem cross-section with 6 mm radius.
Step 1 — Body Weight to Joint Reaction Force
Body weight in Newtons: 80 kg × 9.81 m/s² = 784.8 N. Applying the joint reaction force factor: joint force F = 784.8 × 2.8 ≈ 2,197 N. This is the force transmitted through the hip joint during single-leg stance, substantially higher than the patient's static body weight for the lever-arm reasons covered in the companion joint reaction force article.
Step 2 — Bending Moment at the Stem Cross-Section
The stem's neck geometry creates an offset (moment arm) between the joint reaction force's line of action and the stem's own axis. Converting the 40 mm offset to meters (0.040 m) and computing the moment: M = F × offset = 2,197 × 0.040 ≈ 87.9 N·m. This bending moment is what the stem cross-section has to structurally resist.
Step 3 — Section Modulus of the Circular Cross-Section
For a solid circular cross-section, section modulus Z = πr³/4. Converting the 6 mm radius to meters (0.006 m): Z = π × (0.006)³ / 4 = π × 2.16×10⁻⁷ / 4 ≈ 1.696×10⁻⁷ m³ (equivalently, about 169.6 mm³). Section modulus is a purely geometric property — it describes how efficiently the cross-section's shape resists bending, independent of the material or the applied load.
Step 4 — Bending Stress
Bending stress σ = M / Z = 87.9 / 1.696×10⁻⁷ ≈ 518.3×10⁶ Pa = 518.3 MPa. This is the peak bending stress at the outer fiber of the circular cross-section — the point farthest from the neutral bending axis, where bending stress is always highest for a symmetric cross-section like a solid circle.
Step 5 — Comparing Against a Fatigue Allowable
Using a representative Ti-6Al-4V fatigue strength of 450 MPa: safety margin = 450 / 518.3 ≈ 0.87. A safety margin below 1.0 means the calculated stress exceeds the material's fatigue allowable — a design signal that would prompt increasing the stem cross-section radius, reducing the neck offset, selecting a higher-fatigue-strength material or surface treatment, or re-examining the assumed joint force factor, not proceeding with the design as calculated.
Why Small Geometry Changes Have a Large Effect
Because section modulus for a circular cross-section scales with the cube of radius (Z ∝ r³), a relatively modest increase in stem radius produces a disproportionately large reduction in bending stress — increasing the radius from 6 mm to 7 mm (a 17% increase) increases Z by roughly (7/6)³ ≈ 1.59×, reducing stress by the same factor to about 326 MPa, comfortably below the 450 MPa allowable. This cubic relationship is why implant cross-section sizing is one of the most powerful levers available to a designer working to bring calculated stress within an acceptable margin.