Measurement Uncertainty: Combining Independent Inputs
Combine independent uncertainties in P = V × I by root-sum-square and compare the result with a worst-case bound.
Controls & Instrumentation · Last reviewed
Given
The power of a load is calculated as P = V × I. The voltage is measured as 230.0 V with a standard uncertainty of 0.5 V. The current is measured as 10.00 A with a standard uncertainty of 0.05 A. The two measurements are independent.
Find
The power, its combined standard uncertainty, the expanded uncertainty for a coverage factor k = 2, and the worst-case bound for comparison.
Solution
Related tools and labs
Calculators, labs and simulators are part of the All-Access Pass.
More worked problems
- 4-20 mA Signal Scaling for a Pressure Transmitter
- Control Valve Sizing: Calculate Cv for a Water Line
- Hydrostatic Level Measurement: Pressure to Tank Level
- Differential-Pressure Flow: Why 50% Signal Is Not 50% Flow
- First-Order Sensor Response: Time Constant and Settling
- PLC Analog Input: Raw Counts to Engineering Units
- Size the Branch Circuit for a 40 hp, 460 V Motor
- Voltage Drop on a 240 V Single-Phase Feeder
Questions
What is a standard uncertainty?
An uncertainty expressed as one standard deviation. Manufacturers often give limits instead, and these must be converted before combining them.
What does k = 2 mean?
A coverage factor of 2 gives an interval that contains the true value with about 95% probability when the distribution is roughly normal.
Does this cover calibration and drift?
Only if those are included in the standard uncertainties of the inputs. A full uncertainty budget lists each source.
Educational material. Examples use simplified, hypothetical inputs and are not a design, a construction document or professional engineering advice. Confirm requirements with the codes and standards adopted where the work is built and with a licensed engineer.